Describe the bug
When using the bolus interface on the Loop Apple Watch app, it is possible to dial in a bolus amount with greater precision (0.025 units) than Omnipod can actually deliver. Issue occurs when selecting a bolus value < 1 unit.
Attach an Issue Report
N/A
To Reproduce
Steps to reproduce the behavior:
- Go to Loop app on Apple Watch
- Tap on Bolus
- Use digital crown on the side of the watch to dial the bolus down to a value < 1 unit.
- Observe dose increments of 0.025 units.
Expected behavior
All possible bolus dose selections will conform to the delivery constraints of the given active pump.
Screenshots

Phone
- Hardware: iPhone XS
- Software: iOS 12.2
- Watch Hardware: Apple Watch Series 4
- Watch OS: 5.2
Loop Version
- Version Number: 1.10.0DEV
- Repo: LoopKit/Loop/omnipod-testing
CGM
- Device: Dexcom G6
- Manager app: Dexcom App
Pump
- Manufacturer: Omnipod
- Model: Eros
Additional context
This is based on my understanding that Omnipod's delivery precision is 0.05u/pulse. Apologies if I've misunderstood something here!
Describe the bug
When using the bolus interface on the Loop Apple Watch app, it is possible to dial in a bolus amount with greater precision (0.025 units) than Omnipod can actually deliver. Issue occurs when selecting a bolus value < 1 unit.
Attach an Issue Report
N/A
To Reproduce
Steps to reproduce the behavior:
Expected behavior
All possible bolus dose selections will conform to the delivery constraints of the given active pump.
Screenshots

Phone
Loop Version
CGM
Pump
Additional context
This is based on my understanding that Omnipod's delivery precision is 0.05u/pulse. Apologies if I've misunderstood something here!