Skip to content

Running average on the pressure sensor temperature #48

@TilmanJimenez

Description

@TilmanJimenez

The SBE Data Processing Manual says:

Pressure temperature is computed using a backward-looking, 30-second running average, to prevent bit transitions in pressure temperature from causing small jumps in computed pressure. Because the heavily insulated pressure sensor has a thermal time constant on the order of one hour, the 30-second average does not significantly alter the computed pressure temperature.

but it seems that ctdproc doesn't do this. Is there a specific reason for this?

Metadata

Metadata

Assignees

No one assigned

    Labels

    No labels
    No labels

    Projects

    No projects

    Milestone

    No milestone

    Relationships

    None yet

    Development

    No branches or pull requests

    Issue actions